DEC infant

OVERVIEW

DEC Infant - a disposable EEG cap with 12 Ag/AgCl electrodes.

DEC infant is intended for non-invasive neonatal electroencephalogram (EEG) recording according to international 10-20 system: Fp1, Fp2, 01, 02, C3, C4, T3, T4, Fz, Cz, Pz, GND. The number of electrodes and their arrangement correspond to the recommendations of the IFCN for recording the EEG of newborns. Available sizes: XS (36-42 cm), Inf I (32-36 cm), Inf II (28-32 cm), 
Inf III (24-28 cm).

Advantages:

  • 100% solution to the problem of cross-infection
  • Easy and quick set up - less than 5 min
  • Light and comfortable cap for patient

SPECIFICATION

Number of EEG electrodes 12
Electrode positions FP1, FP2, O1, O2, C3, C4, T7(T3), T8(T4), FZ, CZ, PZ, GND(@AFz)
Electrode polarization < 15 mV
Electrode noise < 2 uV p-p in the passband of 1 Hz to 75 Hz
Electrode impedance less 1 kOhm in the passband of 0.15 Hz to 75 Hz
Material of electrode conductive surface Ag/AgCl
Diameter of the sensor surface 10 mm
Sterility non sterile
Total length of lead cables (product, adapter cable, extension cable) Not more 5.0 m
Electrode condition without paste, need to add EEG conductive paste by self
Avaliable sizes XS (36‑42 cm), Inf I (32‑36 cm), Inf II (28‑32 cm), Inf III (24‑28 cm)
Set up time 3-5 minutes
Storage of the opened package 1 mounth
Amount in a package 5 pcs.
Net weight 3 ± 1 g

Save your time with DEC

  • no need to place of each electrodes
  • no need preparation and special skills
  • no need clean & dry
  • no need disinfection

ACCESSORIES

DEC Infant is connected to amplifier via a special reusable adapters. Adapter DEC Infant ‑DB25 for connection DEC Infant with 12 electrodes to EEG amplifier with common connector DB‑25F. Additional adapter TP2DB25F‑DEC12 for connection EEG cap with common connector DB‑25M to any EEG amplifier via TouchProof 1.5 mm connectors.