FC-NVX2

OVERVIEW

FC-NVX2 functional calibrator intended use for calibration and testing next EEG amplifiers:

• NVX24, NVX36, NVX52

• NVX36T

• NVX136

• actiCHamp Plus

• another with compatible connectors and pinouts.

FC-NVX2 consist of main blocks:

1. MODE selector.

2. The precision voltage dividers with selectable divider ratio of input signals for check of EEG channels. Input signal from external voltage generator after resistive divides connect between all channels EEG and EEG GND lines (MODE in PASSIVE EEG position). For actiCHamp Plus amplifier after the divider can be used low noise and low offset active buffer for decrease output impedance (MODE in ACTIVE EEG position). This is buffer powered from EEG GND 4-pin connector lines.

3. The precision voltage dividers with selectable divider ratio of input signals for check of Auxiliary (AUX). Input signal from external voltage generator after resistive divides connect between AUX and AUX GND connectors or between signal lines of 5 pin AUX connector (MODE in any position).

4. The precision imitator of inter-electrode impedance for check of impedance measuring mode (MODE in IMPEDANCE position).

SPECIFICATION

Parameter Value
Mode Passive EEG Active EEG Impedance
EEG dividers 1:1, 1:10 (10,07), 1:100 (100,8), 1:1000 (999), 1:10000 (9981), Short circuit
EEG divider tolerance ±0,5%
EEG impedance imitation 1 kOhm, 10 kOhm, 20 kOhm, 50 kOhm, 120 kOhm, 400 kOhm
AUX divider tolerance ±0,2%
AUX dividers 1:1, 1:10 (10,07), 1:100 (100,8), 1:1000 (999), 1:10000 (9981), Short circuit
AUX divider tolerance ±0,5%
Input signal max. value ±32 V
Input signal connector BNC
EEG connector type for actiChamp Plus / NVX36 / 36T / 128 KEL40 pins (8830E-040-170S)
EEG connector type for NVX52 KEL50 pins (8830E-050-170S)
GND actiChamp Plus / NVX128 Binder 719, 4 pins (09-9792-20-04)
EEG connector type for NVX24/36 D‑SUB DB‑25F
GND, EEG, AUX+, AUX- connectors type TouchProof 1.5 mm, DIN 42 802‑ST
AUX connector type Binder 719, 5 pins (09-9792-20-05)
LED indicators power from AUX, EEG, EEG GND with Vpp=5 V
Active EEG mode amplifier power from EEG GND with Vpp=5 V NOTE: If Vpp=3 V, then use Uin ≤ 400 mV for Mode Active EEG 1:1.
Case dimensions 210 x 155 x 55 mm
Net weight 0.5 kg

SET INCLUDES

  • Functional calibrator FC-NVX2;
  • NEUROvisor montage imitator NX400031;
  • Cables (DB25-DB25, KEL40-KEL40, KEL40-5KEL40, KEL50-KEL50, GND-AH, AUX-4AUX, AUX-8AUX, TP-TP, 2TP-2TP, SYNC In/out, DB9-DB9, TriggDB9, BNC-2.5 jack);
  • Signal generator NEUROTEST;
  • User manual;
  • Packaging (plastic case).