NVX 72/136/272

OVERVIEW

UNCOMPROMISING DEVICE FOR EEG ACQUISITION

NVX-72/136/272 are designed for a medical-biologic researching in scientific and educational institutions.

Each channel has a direct current input and individual 24-bit analog-to-digital converter for measuring the EEG signal up to 100 thousand times per second.

Possible to extend up to 272 channels by connecting two amplifiers to the media converter, which gives a single clock frequency for all ADCs and ensuring synchronous channels conversion.

Software for planning of experiment and recording signals to EDF+ 16bit, BDF+ 24bit, GDF 32bit.

SOFTWARE

EEG amplifier NVX – 72 / 136 / 272 supports software NeoRec.

NeoRec is a PC software designed for experiment plannig and for recording EEG and other biomedical signals during scientific or medical research, include Evoked Potentials preprocessing.

Record files can be saved in different formats (EDF+ 16 bit, BDF+ 24 bit, GDF 32 bit) or transmited online via stream LSL (Lab Streaming Layer) for analyze by third party software as MATLAB / EEGLAB, OpenViBEetc.

SDK for software development for PC and mobile applications is provided by the manufacturer upon request.

SPECIFICATION

Parameter Value
EEG amplifier
EEG channels 128
Channels of sensors 8 monopolar (4 connectors x 2 channels) or 4 bipolar (4 connectors x 1 channels) set by applied software
Display none
Expansion of channels number up to 408 channels by connecting one or two additional amplifiers to the media converter
Input dinamic range of EEG cannels not less then ± 400 mV
Input impedance more than 1 GOM (with shunting capacity not more than 30 pF)
Relative voltage measurement deviation less ±1%
Input dinamic range of sensor channels not less ±2000 mV set by applied software
Noise of EEG cannels less than 0.9 µV peak-to-peak in the frequency range from 0.1 to 30 Hz
Nonlinearity frequency response of the amplifier from -10% to + 5% (in the frequency range from 0.5 Hz to 100 Hz)
Digitalization 24 bit, 6th order delta-sigma modulator with 64x oversampling, one converter per each channel
Sampling rate (with the corresponding upper frequency at -3 dB) 100 kHz (@16 kHz) (only for 64 EEG channels), 50 kHz (@16 kHz), 25 kHz (9 kHz), 10 kHz (4.9 kHz), 5 kHz (1.6 kHz), 2 kHz (650 Hz), 1000 Hz (300 Hz), 500 Hz (160 Hz), 250 Hz (80 Hz)
Lower cutoff frequency at -3 dB 0..100 Hz or selected from the series 0, 0.001, 0.05, 0.1, 0.5, 1, 2, 5, 10 Hz set by applied software set individually for each channel
Upper cutoff frequency at -3 dB 5000..1 Hz or selected from the series 500, 300, 100, 70, 50, 30, 20, 10, 5, 1 Hz set by applied software set individually for each channel
Electrode impedance measurement range 1–120 kOhm (±10%) @ 30 Hz
Amplifier power From battery module 6 V, 8 A H
Consumption current of amplifier no more than 1400 mA with a data output frequency of 25 000 ... 100 000 Hz; no more than 500 mA with other frequencies of data output; less than 3mA in standby
Operating time from fully charged battery module not less than 4 hours in the modes 25, 50, 100 kHz and not less than 16 hours in the mode 10 kHz
Battery charging time no more than 5 hours
Operating time at indication of the battery status as discharged at least 30 minutes at +25°C
Mode indication Mode indicator
Type of common connector for EEG cap KEL40M (KEL 8830E‑040)
Powering of sensors from amplifier, 5 VDC (±5%) up to 15 mA per probe with electronic protection, but not more than 100 mA in total for 4 channels
MediaConverter
Synchronization channels 8 / 8 not galvanically isolated from the PC
Type of connector for input/output trigger channels DB‑9M / DB‑9F.
Input TTL levels of synchronization channels low level Uil > +0.8 V, Iil > -2.1 mA high level Uih > +3.5 V, Iih > -0.1 mA
Output TTL levels of synchronization channels low level Uol < +0.4 V, Iol < -1.6 mA high level Uoh > +2.8 V, Ioh > 0.5 mA Installed programmatically by a command from the PC. By default, output trigger channels are High.
Synchronization channel mode synchronously with the issued channel readings, accuracy ±0.5 reading.
Media converter input for external clock synchronization of the ADC amplifiers BNC connector characteristic impedance is 50 Ohms, sine / square wave sync signal from 10 to 30 MHz
Control and power of mediaconverter from PC via USB 2.0, 3.0 type B interface
Supply voltage of mediaconverter USB +5 V;
Consumption current of mediaconverter up to 400 mA in active mode, less than 3mA in standby
Сable length 3 m
Mode indication of mediaconverter Mode indicator
General
Safety IEC 60601-1:2005 (ed3.1), IEC 60601-2-26:2012 (ed3.0), IEC 60601-1-2:2014 (ed4.0)
Ingress Protection IPX0
Степень потенциального риска применения class 2a
Mean useful life at least 5 years, if average time of the use does not exceed 80 hours per month
Software technical support during the specified product life
Operating conditions* +0 ... +40°C, relative humidity from 40 to 80%
Storage conditions* +5 ... +40°C, relative humidity to 80%
Transportation conditions* -50 ... +50°C, relative humidity to 90% without condensation

SET INCLUDES

  • NVX 72/136/272 DC EEG amplifier,
  • NVX 72/136/272 media converter,
  • NVX 72/136/272 battery module,
  • NVX 72/136/272 charger,
  • optical cable 1 m,
  • optical cable 10 m,
  • NVX 72/136/272 transport suitcase,
  • Electrode cap MCScap PROFESSIONAL 72/136/272 *.
  • * For MR-compatible system - Electrode cap MCScap PROFESSIONAL MR 72/136/272.